Skip Navigation

Europace 2008 10(5):552-555; doi:10.1093/europace/eun083
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Altemose, G. T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Altemose, G. T.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


ICDs

T-wave oversensing and inappropriate shocks: a case report

Komandoor Srivathsan, Luis R. Scott and Gregory T. Altemose*

Mayo Clinic Arizona—Mayo Clinic Hospital, Cardiovascular Diseases, 5777 East Mayo Boulevard, Phoenix, AZ 85054, USA

A 27-year-old male with congenital long QT syndrome, SCN5A mutation experienced recurrent inappropriate exercise-related implantable cardioverter defibrillator (ICD) shocks. This device showed T-wave oversensing with double, which lead to these device discharges. Dynamic T-wave oversensing was reproducibly provoked at exercise treadmill testing and was confirmed as the mechanism leading to double counting. The insertion of a new pacing and sensing lead with increased R-wave amplitude did not solve the problem. Exchanging the existing ICD generator with one capable of automatic sensitivity control (Biotronik, Lexos DR, Biotronik, Berlin, Germany) completely eliminated T-wave oversensing and inappropriate shocks.

Key Words: Long QT syndrome, T-wave oversensing, Inappropriate shocks


* Corresponding author. Tel: +1 480 342 0239; fax: +1 480 342 1606. E-mail address: altemose.gregory{at}mayo.edu

Manuscript submitted 21 January 2008. Accepted after revision 18 March 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.