T-wave oversensing and inappropriate shocks: a case report
Mayo Clinic Arizona—Mayo Clinic Hospital, Cardiovascular Diseases, 5777 East Mayo Boulevard, Phoenix, AZ 85054, USA
A 27-year-old male with congenital long QT syndrome, SCN5A mutation experienced recurrent inappropriate exercise-related implantable cardioverter defibrillator (ICD) shocks. This device showed T-wave oversensing with double, which lead to these device discharges. Dynamic T-wave oversensing was reproducibly provoked at exercise treadmill testing and was confirmed as the mechanism leading to double counting. The insertion of a new pacing and sensing lead with increased R-wave amplitude did not solve the problem. Exchanging the existing ICD generator with one capable of automatic sensitivity control (Biotronik, Lexos DR, Biotronik, Berlin, Germany) completely eliminated T-wave oversensing and inappropriate shocks.
Key Words: Long QT syndrome, T-wave oversensing, Inappropriate shocks