Europace Advance Access originally published online on May 19, 2008
Europace 2008 10(7):801-808; doi:10.1093/europace/eun125
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrowOA All Versions of this Article:
10/7/801    most recent
eun125v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Google Scholar
Right arrow Articles by Delnoy, P. P.
Right arrow Articles by Plicchi, G.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Delnoy, P. P.
Right arrow Articles by Plicchi, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org
The online version of this article has been published under an open access model. Users are entitled to use, reproduce, disseminate, or display the open access version of this article for non-commercial purposes provided that the original authorship is properly and fully attributed; the Journal, Learned Society and Oxford University Press are attributed as the original place of publication with correct citation details given; if an article is subsequently reproduced or disseminated not in its entirety but only in part or as a derivative work this must be clearly indicated. For commercial re-use, please contact journals.permissions©oxfordjournals.org.


Cardiac resynchronisation therapy

Validation of a peak endocardial acceleration-based algorithm to optimize cardiac resynchronization: early clinical results

Peter Paul Delnoy1,*, Emanuela Marcelli2, Henk Oudeluttikhuis1, Deborah Nicastia1, Fabrizio Renesto3, Laura Cercenelli2 and Gianni Plicchi2

1 Isala Klinieken, Heart Lung Centre, Zwolle, The Netherlands; 2 Clinical Radiology Department, Biomedical Technology Unit, University of Bologna, Bologna, Italy; 3 Clinic Research Department, Sorin Group CRM, Saluggia, Italy

Manuscript submitted 6 February 2008. Accepted after revision 21 April 2008.

* Corresponding author. Tel: +31 38 4242374; Fax: +31 38 4243733. E-mail address: p.p.h.m.delnoy{at}isala.nl


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Aims: Cardiac resynchronization therapy (CRT) involves time-consuming procedures to achieve an optimal programming of the system, at implant as well as during follow-up, when remodelling occurs. A device equipped with an implantable sensor able to measure peak endocardial acceleration (PEA) has been recently developed to monitor cardiac function and to guide CRT programming. During scanning of the atrioventricular delay (AVD), PEA reflects both left ventricle (LV) contractility (LV dP/dtmax) and transmitral flow. A new CRT optimization algorithm, based on recording of PEA (PEAarea method) was developed, and compared with measurements of LV dP/dtmax, to identify an optimal CRT configuration.

Methods and results: We studied 15 patients in New York Heart Association classes II–IV and with a QRS duration >130 ms, who had undergone implantation of a biventricular (BiV) pulse generator connected to a right ventricular (RV) PEA sensor. At a mean of 39 ± 15 days after implantation of the CRT system, the patients underwent cardiac catheterization. During single-chamber LV or during BiV stimulation, with initial RV or LV stimulation, and at settings of interventricular intervals between 0 and 40 ms, the AVD was scanned between 60 and 220 ms, while LV dP/dtmax and PEA were measured. The area of PEA curve (PEAarea method) was estimated as the average of PEA values measured during AVD scanning. A ≥10% increase in LV dP/dtmax was observed in 12 of 15 patients (80%), who were classified as responders to CRT. In nine of 12 responders (75%), the optimal pacing configuration identified by the PEAarea method was associated with the greatest LV dP/dtmax.

Conclusion: The concordance of the PEAarea method with measurements of LV dP/dtmax suggests that this new, operator-independent algorithm is a reliable means of CRT optimization.

Key Words: Heart failure, Cardiac resynchronization therapy, Contractility, Peak endocardial acceleration


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Cardiac resynchronization therapy (CRT) is an established therapy for patients with advanced, congestive heart failure (CHF) associated with cardiac dyssynchrony. The mechanisms of action of CRT are complex, as it interacts with several electromechanical variables, including atrioventricular (AV), interventricular (VV), intraventricular, and intramural delays. Information regarding the optimal programming and follow-up of these devices are limited.

Left ventricular (LV) dP/dtmax and several non-invasive indices have been used to optimize the functions of CRT devices, in particular AV delay (AVD) and VV configuration,1Go–6Go and recent clinical experience has confirmed the need for, and benefits of their long-term reprogramming, as their optimal stimulation settings might change over time.7Go–10Go

A direct assessment of changes in systolic function (LV dP/dtmax) determined to CRT requires invasive measurements, which are neither practical, nor ethically acceptable as a routine procedure.

As a consequence, in clinical practice, the optimization of CRT is usually achieved with the assistance of non-invasive echocardiographic procedures, which can only be performed with the patient supine, and are costly, time-consuming, and operator-dependent.

A haemodynamic sensor integrated in a CRT pulse generator, which automatically calculates and optimizes the programmable settings of the device, might be an objective means of assessing and adjusting therapy, also during activities or exercise.

As the myocardium contracts isometrically, it generates muscular and valve vibrations (endocardial accelerations) that are transmitted throughout the heart and thus measured with an implantable microaccelerometer located inside the tip of a conventional unipolar pacing lead. These vibrations, in their audible component, are responsible for the first heart sound. A pacing system equipped with a microaccelerometer sensor (PEA sensor), able to measure the peak of the endocardial acceleration during the isovolumic contraction phase (first component of the signal here defined as ‘PEA’), has been previously developed and clinically evaluated.11Go,12Go The system allows monitoring of myocardial function by means of PEA, which was identified as an expression of cardiac contractility.13Go Till now, several clinical applications and tools based on PEA assessment have been proposed.14Go–17Go

A new CRT-P device (The New LivingTM CHF, Sorin Group CRM, Saluggia, Italy) equipped with an implantable PEA sensor has been recently developed to monitor cardiac function and guide CRT programming in patients affected by CHF.

This study examined the performance of a new CRT optimization algorithm based on PEA measurements (PEAarea method) in the identification of the optimal CRT pacing configuration, compared with direct measurements of LV pressure (LV dP/dtmax).


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Patient population
The study included 15 patients (mean age = 72 ± 8 years, 13 men) with indications of CRT in accordance with standard international guidance. They were in New York Heart Association (NYHA) CHF functional class II–IV, despite stable optimal medical therapy, including β-adrenergic blockers (n = 14) and angiotensin-converting enzyme-inhibitors or angiotensin II blockers (n = 13). All patients were in sinus rhythm. Their mean echocardiographic LV ejection fraction (EF) was ≤35% (mean = 23 ± 8%) and QRS duration >130 ms (mean = 170 ± 16 ms). The underlying heart disease was ischaemic in eight and non-ischaemic in seven patients. Exclusion criteria were: possible surgical treatment (myocardial revascularization or valve replacement), life-threatening ventricular arrhythmias, right ventricular (RV) dimension superior to LV dimension, myocardial infarction, coronary artery bypass grafting (CABG) within last 3 months, patient already implanted with a conventional pacemaker. The study protocol was approved by our institutional review committee, and all patients granted their informed consent before enrolment.

Implantation of the cardiac resynchronization system
All patients underwent implantation of LivingTM CHF or NewLivingTM CHF CRT pulse generators (Sorin Group CRM), connected to standard atrial and LV pacing leads. The LV lead was inserted transvenously through the coronary sinus, preferably into a lateral or postero-lateral cardiac vein. A pacing lead (BESTTM, Sorin Group CRM) with a microaccelerometer sensor inside its tip was implanted at the RV apex to record the PEA signal. At implant, all pulse generators were programmed in dual chamber atrioventricular synchronous (DDD) mode and simultaneous biventricular (BiV) stimulation (VV = 0), and the AVD was adjusted to ensure full electrical pre-excitation of the ventricles during DDD pacing. The average AVD was 154 ± 20 ms.

Catheterization procedure
At a mean interval of 39 ± 15 days after implantation of the CRT system, the patients underwent cardiac catheterization under mild sedation. Haemodynamic measurements were performed using a 7F pigtail catheter equipped with a high-fidelity pressure transducer (CD-Leycom, Zoetermeer, The Netherlands). The catheter was advanced into the LV from the left femoral artery and connected to a signal-conditioning module (Sentron, Roden, The Netherlands) to record pressure.

Stimulation protocol
After placement of the pressure catheter, the heart was stimulated programming the device with a PMP 2000T external programmer (Sorin Group CRM).

The protocol included single LV and BiV stimulation, either simultaneously (BiV0), or sequentially at VV intervals of 12 or 40 ms, with LV (LR12, LR40) or RV (RL12, RL40) preactivation, in a random order. At each of these configurations, the AVD was scanned between the shortest 60 ms and the longest 220 ms interval, with four increments of AVDstep calculated as (PR – 90)/4 ms. The five tested AVD intervals were: 60 ms, 60 + 1 x AVDstep ms, 60 + 2 x AVDstep ms, 60 + 3 x AVDstep ms, and 60 + 4 x AVDstep ms. The AVD scans performed for each patient are listed in Table 1.


View this table:
[in this window]
[in a new window]

 
Table 1 Scans of the atrioventricular delay

 
Data sampling and analysis
PEA signals, LV pressure with LV dP/dtmax calculation, and analog electrocardiographic signals were recorded simultaneously, using a MP100 acquisition system (Biopac System Inc., Goleta, CA, USA) at a sampling rate of 1000 Hz for each channel. For each pacing configuration, a 10-s programming period was required and a 30-s recording period was preceded by a stabilization period of 30 s (i.e. a total of 70 s for each configuration). According to the stimulation protocol, six pacing configurations were tested (RL40, RL12, BiV0, LR12, LR40, LV), performing AVD scanning over five AVD intervals. Thus, the total time needed to find the optimal pacing configuration with the new PEAarea method was approximately 35 min (2100 s), calculated as (70 s) x (6) x (5) = 2100 s.

For PEA recordings, the voltage signal generated by the acceleration was measured from peak to peak in gravity (1 g = 9.8 m/s2). LV dP/dtmax and PEA were measured at steady state, cycle-by-cycle, using the AcqKnowledge 3.8.1 analysis software (Biopac System Inc., Goleta, CA, USA), and values, sampled over approximately 30 s, were averaged. Cycles with failure to capture, affected by extrasystolic events and post-extrasystolic cycles were excluded from the analysis. LV dP/dtmax and PEA were first measured during intrinsic rhythm, used as the reference patient's status condition. The heart rate was kept constant by atrial pacing for all subsequent stimulation sequences.

Optimization of cardiac resynchronization
Response to CRT was defined as a ≥10% increase in LV dP/dtmax, considered the gold standard to monitor cardiac function in clinical CRT studies,9Go,10Go as it is reliable, reproducible, and not operator-dependent. For each patient, the optimal CRT configuration was determined by comparing the optimization criteria based on measurements of LV dP/dtmax (LV dP/dtmax method) vs. measurements of PEA (PEAarea method).

The new PEAarea method
Studies in animals have shown that the peak-to-peak value of acceleration of heart vibrations (PEA), recorded during LV isovolumic contraction with an accelerometer sensor, corresponds to the endocardial recording of the first heart sound (S1).18Go LV contractility, as expressed by the maximum rate of rise of LV pressure (LV dP/dtmax), and the respective timing of atrial and ventricular systoles (the AVD), are both major and independent determinants of the amplitudes of S1 and PEA, in normal and in failing hearts.19Go–24Go In normal hearts, shortening of the AVD causes a rapid increase in PEA amplitude, mainly determined by the degree of opening of the mitral valve leaflets, strictly related to an increase in transmitral flow during the A wave, while LV dP/dtmax remains relatively fixed. Similar results have been shown in clinical studies of CRT, where LV dP/dtmax remained nearly fixed with shortening of the AVD in case of successful resynchronization procedures, whereas it decreased with shortening of the AVD in case of unsuccessful resynchronization attempts.10Go

The new PEAarea method is, according to these results, based on measurements of the area under the PEA curve: it is expected to have a lower PEAarea in case of unsuccessful resynchronization attempts, because of LV dP/dtmax decreasing for short AVDs, with respect to successful resynchronization procedure in which LV dP/dtmax remained nearly fixed. For each tested VV interval setting, the area of PEA curve was estimated, using a simplified formula, as the average of PEA values measured during AVD scanning (Figure 1A–D). Thus, the PEAarea is an index of both contractility variations and transmitral flow variations, and we assumed that an increase in the area reflects an increase in overall haemodynamic benefit achieved by the programmed configuration.


Figure 1
View larger version (31K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 (AD) Representative examples among cardiac resynchronization therapy responder patients of (peak endocardial acceleration) PEAarea curve measured during atrioventricular delay scanning for different pacing configurations [RL40: sequential biventricular pacing with 40 ms of right ventricular (RV) preactivation; RL12: sequential biventricular pacing with 12 ms of RV preactivation; LR12: sequential biventricular pacing with 12 ms of left ventricular (LV) preactivation; LR40: sequential biventricular pacing with 40 ms of LV preactivation; BiV0: simultaneous biventricular pacing].

 
Statistical analysis
The mean values of LV dP/dtmax, calculated over a period of 10 respiratory cycles at each pacing configuration, were compared, using the Student-Newman-Keuls test for comparison of multiple variables. This specific t-test for comparison of paired variables belonging to a group, takes into account, for each paired comparison, the influence of all variables included in the group. A value of {alpha}{tau} < 0.05 was considered statistically significant.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Responder patients were identified by a ≥10% increase in LV dP/dtmax. Among the 15 patients included in this study, 12 (80%) were responders, a proportion consistent with other clinical studies of CRT.9Go,10Go In 9 out of the 12 responders (75%), the results of the LV dP/dtmax and PEAarea methods were concordant (Table 2). The haemodynamic responses (LV dP/dtmax and PEAarea trends) for different pacing configurations are shown in Figure 2A–D for the same four representative patients used for showing the PEAarea curve during AVD scanning in Figure 1A–D.


Figure 2
View larger version (13K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 (A–D) Comparison between (peak endocardial acceleration) PEAarea method and LV dP/dtmax method in identifying the optimal stimulation configuration for the same representative responder patients reported in Figure 1. Measurements made during intrinsic rhythm were used as baseline to calculate % variation of PEAarea and LV dP/dtmax indexes. Symbol # indicates absence of significant difference between the two or more cardiac resynchronization therapy configurations by Student-Newman-Keuls analysis.

 


View this table:
[in this window]
[in a new window]

 
Table 2 Concordance of the optimal pacing configuration identified by the LV dP/dtmax vs. PEAarea methods

 
The overall means and standard deviations for LV dP/dtmax and PEAarea values for pacing configuration for each patient were reported in Table 3.


View this table:
[in this window]
[in a new window]

 
Table 3 Means and standard deviation (SD) for LV dP/dtmax and PEAarea values for each pacing configuration, for each patient

 
Both methods identified the greatest haemodynamic improvement with LV preactivation or single-chamber LV stimulation in six, RV preactivation in three, and simultaneous biventricular stimulation in two responders. LV preactivation or single chamber LV stimulation was associated with the greatest haemodynamic improvement by at least one method in nine, RV preactivation in five, and BiV stimulation in four responders. Finally, in eight responders, more than one optimal stimulation configuration (range 2–4) were identified by the LV dP/dtmax method (including opposite ventricular preactivation in two patients), whereas the PEAarea method identified a single optimal configuration in all responders (Table 2).

For each patient, the values of LV dP/dtmax increase in % for the optimal pacing mode, compared with PEAarea variations in % for the optimal pacing mode, were summarized in a scatter diagram plot (Figure 3).


Figure 3
View larger version (15K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 Scatter diagram plot of LV dP/dtmax % variations vs. (peak endocardial acceleration) PEAarea % variations, for each patient in the optimal pacing configuration.

 

    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
With the currently available CRT pulse generators, the AVD and the VV interval have become programmable, and their optimization is generally recommended, as an appropriate selection of VV interval might further optimize LV function, at implant as well as during follow-up.25Go–32Go Therefore, a direct comparison of a new method of CRT optimization, based on PEA measurements, with LV dP/dtmax, the haemodynamic gold standard, was of particular interest. LV dP/dtmax has been used previously as an index of LV performance in CRT.27Go–29Go However, it is influenced by heart rate, preload, LV synchronization, and degree of mitral regurgitation. Furthermore, when applied to optimize CRT, it requires an invasive procedure, and its value is limited to short-term observations. On the contrary, the optimization of CRT based on individual adjustments of heart rate, AVD, VV interval, and mode of stimulation can be accomplished non-invasively in the long term, with the assistance of haemodynamic sensors integrated in the implanted devices. The perspective of being able to re-adjust CRT, with an automatic on board function for periodic CRT optimization by PEA, at rest and during lifetime exercise activity, is attractive and preferable to the echocardiographic procedures that are currently applied in clinical practice only at follow-up. Moreover, the echocardiographic method is not so sensitive and might not be able to detect subtle changes in LV systolic function. Therefore, the consistent results obtained with the PEAarea method, validated by the direct measurements of LV dP/dtmax, strongly support the introduction of this new operator-independent method of CRT optimization into clinical practice and implantable devices.

Study limitations
The absence of an automated stimulation protocol and data analysis system limited the number of AVD used to calculate the area of PEA curve and VV intervals we were able to test. However, we believe that for preliminary evaluation of the new PEA-based method a five-step AVD scanning may be enough to provide meaningful data to estimate area under the PEA curve.

The number of patients included in the present study was relatively small, even if the rate of CRT responders was comparable with data reported in literature.33Go

Furthermore, our method, based on the measurement of PEA during scanning of the AVD, is not applicable to patients presenting with atrial fibrillation.

Among 12 responders, according to the LV dP/dtmax measurement, three patients showed no concordant results between LV dP/dtmax and PEAarea method indications. For one patient, during the procedure, we encountered some problems with the pressure catheter that required its repositioning. Even if we used the same calibrating procedure, the obtained LV dP/dtmax trend could have been altered because of LVP recordings assessed before and after catheter repositioning.

For the other two patients, the optimal CRT configuration indicated by PEAarea method is ‘adjacent’ to the best configuration indicated by LV dP/dtmax. This means that the optimal configuration indicated by LV dP/dtmax corresponds to the ‘second optimal’ configuration indicated by PEAarea slightly lower than the optimal PEAarea value. Small differences can be observed between the optimal and the ‘second optimal’ CRT configurations indicated by PEAarea; consequently, for these two patients we can say that, even if PEAarea and LV dP/dtmax methods does not provide exactly the same CRT best mode, however it is clear that both identified the same optimal range of CRT configurations (RV preactivation—RL12, RL40—in one case, and LV preactivation—LR12, LR40—for the other). Moreover, for sequential BiV stimulation we performed a VV interval scanning with only two increments of VV step (12 ms, 40 ms). Probably, a more refined VV interval scanning (e.g. four increments of VV step, from 12 to 40 ms) would have helped in determining the optimal configuration for these two ‘no concordant’ patients, as for both LV dP/dtmax and PEAarea methods the best pacing mode could correspond to a configuration with VV interval included in the range of 12–40 ms.

We are working towards implementing the CRT-P device as an automated stimulation protocol and PEAarea calculations during AVD scanning in order to provide a more practical and systematic method to apply for a more large and complete clinical study. With this automated PEAarea algorithm implemented in the device, the time needed to find the optimal pacing configuration can be reduced up to 10–15 min, minimizing the programming and the recording periods.

Finally, the outcomes of CRT optimization based on the PEAarea method might not be predicted over an intermediate or long-term follow-up. Thus, the long-term performance of the PEAarea index must be clinically validated.

In the next clinical study, we are planning to validate the method; results obtained with the automated PEAarea algorithm implemented in the device will be compared with clinical non-invasive measurements of cardiac performance and CRT outcome (e.g. the 6 min-walk test), at implant and during periodic follow-up (three or six months after implant).


    Conclusions
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Recent clinical studies have shown that CRT might require reprogramming during follow-up. The proposed method, based on measurements of the area under the PEA curve during scanning of the AVD with different CRT configurations, might be the answer to the need of automatic CRT optimization at implant and during follow-up. The attainment of concordant results with invasive haemodynamic measurements in the majority of responders warrant further testing of this operator-independent and expeditious method of CRT optimization in a larger patient population.


    Funding
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
Funding to pay the Open Access publication charges for this article was provided by Sorin Group.

Conflict of interest: none declared.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Funding
 References
 
[1] Eugene M, Lascault G, Frank R, Fontaine G, Grosgogeat Y, Teillac A. Assessment of the optimal atrio-ventricular delay in DDD paced patients by impedance plethysmography. Eur Heart J (1989) 10:250–5.[Abstract/Free Full Text]

[2] Kindermann M, Fröhlig G, Doerr T, Schieffer H. Optimizing the AV delay in DDD pacemaker patients with high degree AV block: mitral valve Doppler versus impedance cardiography. Pacing Clin Electrophysiol (1997) 20:2453–62.[CrossRef][Medline]

[3] Ovsyshcher IE. Toward physiological pacing: optimization of cardiac hemodynamics by AV delay adjustment. Pacing Clin Electrophysiol (1997) 20:861–5.[CrossRef][Medline]

[4] Ishikawa T, Sumita S, Kimura K, Kikuchi M, Kosuge M, Kuji N, et al. Prediction of optimal atrioventricular delay in patients with implanted DDD pacemakers. Pacing Clin Electrophysiol (1999) 22:1365–71.[CrossRef][Medline]

[5] Ritter P, Padeletti L, Gillio-Meina L, Gaggini G. Determination of the optimal atrioventricular delay in DDD pacing. Comparison between echo and peak endocardial acceleration measurements. Europace (1999) 1:126–30.[Abstract/Free Full Text]

[6] Meluzin J, Novak M, Mullerova J, Krejci J, Hude P, Eisenberger M, et al. A fast and simple echocardiographic method of determination of the optimal atrioventricular delay in patients after biventricular stimulation. Pacing Clin Electrophysiol (2004) 27:58–64.[CrossRef][Medline]

[7] O'Donnell D, Nadurata V, Hamer A, Kertes P, Mohammed W. Long-term variations in optimal programming of cardiac resynchronization therapy devices. Pacing Clin Electrophysiol (2005) 28:S24–S26.[CrossRef][Medline]

[8] De Cock CC, Van Campen LM, Jessurun ER, Allaart CA, Vos DS, Visser CA. Long-term follow-up of patients with refractory heart failure and myocardial ischemia treated with cardiac resynchronization therapy. Pacing Clin Electrophysiol (2005) 28:S8–S10.[CrossRef][Medline]

[9] Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, et al, Pacing Therapies in Congestive Heart Failure (PATH-CHF) Study Group. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol (2002) 39:2026–33.[Abstract/Free Full Text]

[10] Auricchio A, Stellbrink C, Block M, Sack S, Vogt J, Bakker P, et al. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. Circulation (1999) 99:2993–3001.[Abstract/Free Full Text]

[11] Clementy J. Dual chamber rate responsive pacing system driven by contractility: final assessment after 1-year follow-up. The European PEA Clinical Investigation Group. Pacing Clin Electrophysiol (1998) 21:2192–7.[CrossRef][Medline]

[12] Langenfeld H, Krein A, Kirstein M, Binner L. Peak endocardial acceleration-based clinical testing of the ‘BEST’ DDDR pacemaker. European PEA Clinical Investigation Group. Pacing Clin Electrophysiol (1998) 21:2187–91.[CrossRef][Medline]

[13] Rickards AF, Bombardini T, Corbucci G, Plicchi G. An implantable intracardiac accelerometer for monitoring myocardial contractility. The Multicenter PEA Study Group. Pacing Clin Electrophysiol (1996) 19:2066–71.[CrossRef][Medline]

[14] Ritter P. Diagnostic functions of implanted sensors. In: Progress in Clinical Pacing—Santini M, ed. (1996) Armonk: New York Futura Media Services Inc. 183–8.

[15] Oude Luttikhuius HA. A new approach for chronicle comparison of apical vs. septal right ventricular stimulation. A case study. In: Pacemaker Leads 1997—Antonioli GE, ed. (1997) Bologna, Italy: Monduzzi. 253.

[16] Deharo JG, Peyre JP, Ritter P, Chalvidan T, Macaluso G, Djiane P, et al. Adaptive rate pacing controlled by a myocardial contractility index for treatment of malignant neurally mediated syncope (abstract). PACE (1997) 20:1192.

[17] Deharo JG, Peyre JP, Ritter P, Chalvidan T, Berland Y, Djiane P. A Sensor-based evaluation of heart contractility in patients with head-up tilt-induced syncope. PACE (1998) 21:223.[Medline]

[18] Wood JC, Festen MP, Lim MJ, Buda AJ, Barry DT. Regional effects of myocardial ischemia on epicardially recorded canine first heart sounds. J Appl Physiol (1994) 76:291–302.[Abstract/Free Full Text]

[19] Stept ME, Heid CE, Shaver JA, Leon DF, Leonard JJ. Effect of altering P-R interval on the amplitude of the first heart sound in the anesthetized dog. Circ Res (1969) 25:255–63.[Abstract/Free Full Text]

[20] Dupuis JM, Kobeissi A, Vitali L, Gaggini G, Merheb M, Rouleau F, et al. Programming optimal atrioventricular delay in dual chamber pacing using peak endocardial acceleration: comparison with a standard echocardiographic procedure. Pacing Clin Electrophysiol (2003) 26:210–3.[CrossRef][Medline]

[21] Bombardini T, Gaggini G, Marcelli E, Parlapiano M, Plicchi G. Peak endocardial acceleration reflects heart contractility also in atrial fibrillation. Pacing Clin Electrophysiol (2000) 23:1381–5.[CrossRef][Medline]

[22] Marcelli E, Plicchi G, Parlapiano M, Gaggini G. Feasibility of detecting sustained ventricular tachycardias with an implantable sensor (abstract). Pacing Clin Electrophysiol (2003) 26:S166.

[23] Marcelli E, Plicchi G, Parlapiano M, Marini S. Epicardial vibrations vector assessed with a triaxial micro-accelerometer in experimental left ventricular dysfunction (abstract). Pacing Clin Electrophysiol (2003) 26:1081.

[24] Plicchi G, Marcelli E, Parlapiano M, Bombardini T. PEA I and PEA II based implantable haemodynamic monitor: pre-clinical studies in sheep. Europace (2002) 4:49–54.[Free Full Text]

[25] Melzer C, Knebel F, Ismer B, Bondke H, Nienaber CA, Baumann G, et al. Influence of the atrio-ventricular delay optimization on the intra left ventricular delay in Cardiac Resynchronization Therapy. Cardiovasc Ultrasound (2006) 4:5.[CrossRef][Medline]

[26] Sogaard P, Egeblad H, Pedersen AK, Kim WY, Kristensen BO, Hansen PS, et al. Sequential versus simultaneous biventricular resynchronization for severe heart failure: evaluation by tissue Doppler imaging. Circulation (2002) 106:2078–84.[Abstract/Free Full Text]

[27] Auricchio A, Ding J, Spinelli JC, Kramer AP, Salo RW, Hoersch W, et al. Cardiac resynchronization therapy restores optimal atrioventricular mechanical timing in heart failure patients with ventricular conduction delay. J Am Coll Cardiol (2002) 39:1163–9.[Abstract/Free Full Text]

[28] Jansen AH, Bracke FA, van Dantzig JM, Meijer A, van der Voort PH, Aarnoudse W, et al. Correlation of echo-Doppler optimization of atrioventricular delay in cardiac resynchronization therapy with invasive hemodynamics in patients with heart failure secondary to ischemic or idiopathic dilated cardiomyopathy. Am J Cardiol (2006) 97:552–7.[CrossRef][ISI][Medline]

[29] Van Gelder BM, Bracke FA, Meijer A, Pijls NH. The hemodynamic effect of intrinsic conduction during left ventricular pacing as compared to biventricular pacing. J Am Coll Cardiol (2005) 46:2305–10.[Abstract/Free Full Text]

[30] Mortensen PT, Sogaard P, Mansour H, Ponsonaille J, Gras D, Lazarus A, et al. Sequential biventricular pacing: evaluation of safety and efficacy. Pacing Clin Electrophysiol (2004) 27:339–45.[CrossRef][Medline]

[31] Van Gelder BM, Bracke FA, Meijer A, Lakerveld LJM, Pijls NHJ. Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapy. Am J Cardiol (2004) 93:1500–3.[CrossRef][ISI][Medline]

[32] Perego GB, Chianca R, Facchini M, Frattola A, Balla E, Zucchi S, et al. Simultaneous vs. sequential biventricular pacing in dilated cardiomyopathy: an acute hemodynamic study. Eur J Heart Fail (2003) 5:305–13.[CrossRef][ISI][Medline]

[33] Peraldo C, Achilli A, Orazi S, Bianchi S, Sassara M, Laurenzi F, et al. Results of the SCART study: selection of candidates for cardiac resynchronisation therapy. J Cardiovasc Med (2007) 8:889–95.[ISI]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrowOA All Versions of this Article:
10/7/801    most recent
eun125v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager